ABR Practice Exam Question 7: Answer
The answer is statement 5 (None of the above).
As per AAPM TG-51, Mraw (one of Mraw+ or Mraw-) in the denominator is the reading corresponding to the charge collected for the reference dosimetry measurements in the clinic and which should be the same as for the chamber calibration.